产品描述 µ-Gard 气体变送器用于连续检测环境中的各种气体,广泛应用于各种环境,性能可靠,连接方便 应用 可适合各种商业应用,基于标准的4–20 mA 或 2-10 V模拟量输出,µ-Gard可配合各种控制器使用 特点 ·连续检测 ·便于维护和校正 ·内置电源保护电路 ·过载和短路保护 ·长寿命传感器 ·低零点漂移 ·有效防止中毒 ·4 – 20 mA 标准输出 ·可选2 - 10 V 输出 技术参数 | MA-0 CO | MA-9 O2 | MA-9 NO | MA-2 NO2 | MA-5 SO2 | MA-2 NH3 | MA-2 Ex | MA-2 NH3 | MA-3 NH3 | MA-4 Freon | MA-4 Ex | MA-3 ETO | 传感器原理 | 电化学,扩散式 | 催化燃烧式 | 半导体 | 响应时间 t90 (s) | ≤ 50 | ≤ 15 | ≤ 45 | ≤ 25 | ≤ 15 | ≤ 120 | ≤ 10 (甲烷) | ≤ 100 | ≤ 50 | ≤ 30 | ≤ 50 | 分辨率 | ±3ppm | ±0.1vol. % | ±0.2 ppm | ±0.2 ppm | 0.2 ppm | <±20 ppm | ± 1% of reading | --- | --- | --- | --- | 重复性 (%) | ± 3 % | --- | --- | ± 2 % | ± 2 % | --- | --- | --- | --- | --- | --- | 零点漂移/年 | < 5% | --- | <0.3 ppm | <0.2 ppm | <0.2 ppm | --- | ±<6%LEL (Methane) | --- | --- | --- | --- | 灵敏度漂移/月 | --- | < 4%/12 | < 5%/12 | < 2%/1 | < 2%/1 | < 2%/1 | ± < 1.5% Methane | --- | --- | --- | --- | 传感器寿命 (正常使用环境)年 | | ≥ 2 | ≥ 2 | ≥ 2 | ≥ 2 | ≥ 2 | ≥ 3 | ≥ 5 | 最长存贮时间 | 6 months | 12 months | 操作环境 | 湿度(无冷凝) | 15 – 90 % RH | 15 – 90 % RH | 15 – 95 % RH | 温度 | 工作温度 | -10 to + 50°C | -30 to + 50 °C | -15 to + 50°C | -15 to + 50 °C | 存贮温度 | 5 C to + 40 °C | 0 to + 50 °C | 0 to + 50 °C | 工作压力 | 1大气压±10% | 规格参数 供电 18 - 28 VDC 耗电量 电化学传感器: 22 mA, 0,6 VA 催化/半导体传感器: 35 mA, max. 0.85 VA 外壳(F) GW 塑料 75 GWT 阻燃性 UL 94: V0 外壳颜色 RAL 7032 (浅灰) 尺寸(d x H) 87 x 45,5 mm 重量 约 0.2 kg 防护等级 IP 43 安装方式 壁挂式 线缆 标准3线 线缆连接 0.25 mm2 (24 AWG)- 2.5 mm2 (14 AWG) 环路负载电阻500 Ω 输出信号 模拟输出信号 标准 4-20 mA 4 – 20 mA, ≤ 500 Ω带过载和短路保护 可选 2-10 V 2 – 10 V, ≥ 50 kΩ带过载和短路保护 执行标准 EMC 2004/108/EEC, CE 质保期 12个月(不含传感器) |